MedQraft
← All insights
QMS / ISO 13485

ISO 13485 7.3.6 design verification: plans, test records and reports

Connect design requirements to verification plans, test records and reports.
Understand review versus verification, document controls, traceability and design and development files.

Last regulatory review: 2026-11-01 Rev. 1.0 13 min read

A verification conclusion needs a requirement and objective evidence

Design verification establishes whether design outputs meet design input requirements through objective evidence.
A label saying “testing complete” does not explain what was demonstrated.
Readers need to identify the requirement, the design version, the method and acceptance criteria, the actual results and the resulting conclusion.

ISO 13485:2016 clause 7.3.6 addresses verification under documented plans.
These include methods, acceptance criteria and, where appropriate, statistical techniques with a rationale for sample size.
Results, conclusions and necessary actions are retained.
MDSAP describes verification evidence from testing, measurement and analysis, with acceptance criteria established before execution.[1]

The document names, identifiers, assignments and examples below are practical suggestions from MedQraft.
They do not mean that ISO requires three separate files named “plan,” “test record” and “report.”
An integrated document can work, provided planned work, actual execution and conclusions remain distinguishable and connected.

Review evaluates the design; verification establishes input conformity

Under 7.3.5, relevant participants assess the ability of design results to meet requirements and identify problems and necessary actions.
Verification under 7.3.6 establishes evidence that outputs meet inputs.
The distinction is not simply “meeting versus physical test.”
Calculation, analysis and examination of design information may also be appropriate verification methods.[1]

Consider a fictional thermometer.
A reviewer identifies a concern: a previous temperature remains on screen when a new measurement fails, potentially appearing to be a new result.
The review produces a finding and actions for revision and evaluation.
Once the input defines the required behavior and the display design is revised, verification checks implementation in the specified error conditions.
If the requirement itself changes, that change also needs appropriate control.

Review leads from design assessment to issues and actions; verification connects inputs and outputs through evidence
Figure 1.
Review and verification have different, connected purposes.
A review may assess verification evidence; agreement at a meeting does not replace the results of the required evaluation.

Verification plan: establish the scope, method and criteria before execution

A plan or protocol aligns what will be checked with how the outcome will be judged.
Organizing it around requirement identifiers helps expose missing coverage.
The following is an example structure, not a prescribed form.

Element What to document and why
Scope and configuration Model, hardware and software versions, drawing and specification revisions.
Define the design to which the results will apply.
Inputs and outputs Requirement identifiers and the relevant output documents and sections.
State what each verification activity addresses.
Method and conditions Initial state, stimulus or input, sequence, observation point and end condition.
Reduce assumptions left to the operator.
Acceptance criteria The expected observation or permitted measurement, its source or rationale and applicable conditions.
Samples and repetitions Selection rationale, representation of product variants, necessary repetition and statistical approach.
Avoid unexplained customary numbers.
Resources and responsibilities Equipment and tools, execution and technical review responsibilities, approval authority, timing and prerequisites under the organization's procedures.

For a display-behavior test, “cause an error” is insufficient.
Specify the error state and how it will be reproduced.
Does the starting configuration contain a previous valid result? Is the error introduced after a new measurement starts? Which area of the display is observed, and at what point? These choices affect interpretation.
Specific timing and limits should come from the product's controlled inputs, outputs and risk analysis; this example does not prescribe universal values.

A developer or test engineer familiar with the requirement may draft the plan.
A technical reviewer can assess whether the method actually addresses the requirement and whether the criteria and samples are justified.
An authorized approver releases it under the organization's procedure.
These are example responsibilities, not mandatory department names or a universal number of signatures.

Where intended use requires connection or an interface with other medical devices, the verification scope also needs to address input conformity in that connected configuration.[1] Evidence from standalone operation should not be presented as evidence for an untested connection.

Test records: distinguish the intended procedure from actual execution

Test records capture what happened during execution.
They should be created when work is performed and observations are made, rather than reconstructed when the report is written.
Copying expected results from the protocol into the actual-results field obscures what was observed.

A useful record structure connects the test item with the date, operator, specimen identity, device and software configuration, equipment, actual conditions and observed or measured results.
Link screenshots, logs and measurement files to the execution identifier and a retrievable storage location.
Where measurement equipment is used, connect the relevant calibration or verification status and suitability to the measurement concerned.

Suppose a screenshot shows that the previous value disappeared, but the associated log identifies a different software version.
Before assigning a pass, establish what was tested and assess the discrepancy's effect on the conclusion.
If execution used a different instrument or condition from the protocol, record the deviation, reason and assessment rather than silently treating it as planned work.

An interrupted test or a condition that could not be reproduced should have an explicit status, such as “not performed” or “conclusion pending,” under the organization's terminology.
Keep missing evidence distinguishable from a demonstrated failure.
Neither supports an unsupported pass.

Verification report: explain requirement-level conclusions and limits

The report connects the controlled plan and actual records to a conclusion about input conformity.
Identify the plan revision, tested configuration, executed scope and source records.
Summarize the comparison against each criterion and the resulting conclusion.
Technical review should check consistency with raw data and the treatment of missing work, deviations and unfavorable findings.
Approval should leave remaining actions and the scope of the conclusion clear.

An external laboratory report also needs an internal connection.
Match its specimen and configuration, method and edition, test scope and findings to your requirement identifiers.
Explain what the report demonstrates.
Attaching the file does not justify extending its conclusion to functions or product versions that were not evaluated.

Within its scope for US premarket submissions, FDA's non-clinical bench performance testing guidance recommends information on methods, samples, acceptance criteria, results and conclusions, including the effect of protocol deviations.[2] Software verification and validation and human factors are excluded from this guidance; it is not the direct test basis for the fictional display software.
This submission guidance should not be turned into a mandatory ISO template for every product.
Internal verification records and a regulatory submission summary serve different purposes and may require different levels of detail.

One requirement identifier connects a verification plan, execution records and raw data, and the result report
Figure 2.
The plan explains method and criteria, the record captures execution and observations, and the report explains the judgment.
Shared requirement and test identifiers make the evidence retrievable across these documents.

Worked documentation example: follow one requirement to its conclusion

The following identifiers and content illustrate a fictional documentation structure.
They are not actual test results.
The display requirement is not a universal requirement for thermometers.

Connection Fictional entry
Input REQ-01 Do not display the previous measurement in the current-result area during the specified measurement-error state.
Reference the input section defining applicable states.
Output UI-02 Identify the revision of the display layout and state-transition specification implementing the requirement.
Plan VP-01 / item T-01 Starting with a previous value stored, reproduce the specified error and observe the current-result area.
The approved protocol defines each condition and observation point.
Record RUN-01 Enter the actual specimen, version, conditions, time and observation; link screen evidence and logs.
Actual results are entered after execution.
Report VR-01 Compare T-01 results with the predetermined criterion to determine REQ-01 conformity.
Reflect unexecuted conditions or deviations in the conclusion's scope.
Design and development file Include or reference REQ-01, UI-02, VP-01, RUN-01, VR-01 and subsequent action and change records.

A traceability matrix can index these connections, but a “pass” cell does not replace raw data.
When one requirement covers several error conditions, retain coverage and status at the appropriate level so that a partial execution does not appear to establish complete conformity.

Which standard controls these documents and records?

ISO 13485:2016 provides the QMS framework for this documentation.
Clause 7.3.6 concerns verification, 4.2.4 and 4.2.5 concern document and record controls, and 7.3.10 concerns design and development files.
Consider 7.3.9 when designs change, and 7.6 when monitoring or measurement resources provide conformity evidence.[1][3] The following are practical ways to connect these controls to the work.

Material Example control implementation
Plan or protocol Identify document number, revision and scope; establish review and approval before issue.
Make the applicable approved version available and track changes and effective dates.
Execution records and raw data Connect execution and specimen identities.
Define access and modification permissions, retrievable storage and protection; keep changes identifiable.
Report Reference the source records and plan revision and identify the reviewed and approved conclusion.
Explain revised conclusions and their impact rather than silently replacing the previous conclusion.
External standards and reports Register the applicable document and edition and the identity of external evidence.
Assess whether revisions affect existing plans or conclusions.
Design and development file Include or reference verification and change records for the device type or family.
Keep references usable when storage locations change.

It is helpful to distinguish a protocol revision from correction of an execution record.
Changing a method changes the procedure to be applied; correcting a record addresses a fact already recorded.
Preserving the original entry together with who corrected it, when and why supports confidence in the result.
In electronic systems, implement access control, backup, change history and long-term readability through procedures and appropriate system functions.
Saving a PDF alone does not establish all these controls.

Set retention in the organization's procedures after assessing document type, device lifetime and applicable market requirements.
MDSAP distinguishes additional jurisdiction-specific retention requirements.[1] Do not prescribe one unexplained retention period for every record or assume that development completion is always the starting date.

Controlled plan revisions and separate execution records remain linked through the report and the design and development file
Figure 3.
Separate plan revisions from execution records and identify the revision actually used by each report.
New protocols and retests do not erase the connection to earlier results and necessary actions.

Select technical test standards separately from QMS control clauses

ISO 13485 does not prescribe one display time, accuracy limit or sample count for every thermometer.
Select technical standards according to intended use, measurement principle, functions and target market, then establish the applicable edition, clauses and scope.
A standard number alone is less useful than a connection to the actual test item, criterion and any excluded or modified conditions and their rationale.

This fictional example addresses one display behavior, so it does not declare a specific thermometer test standard applicable to the entire product.
In an actual project, requirements related to software, usability or risk controls should connect to the relevant activity plans and applicable standards.
Demonstrating that the previous value disappears does not automatically demonstrate that intended users correctly understand the error.
Validation of the specified application or intended use has the distinct purpose addressed by 7.3.7.

Preserve the evidence chain after failures and changes

When a result misses its criterion, retain both the observation and the criterion, then assess causes, impact and necessary action.
Relaxing a criterion after execution merely to obtain a pass undermines the original judgment.
If the criterion itself is wrong, evaluate the justification for its change, the impact on requirements, risk and previous results, and the work needed to establish a new conclusion.

For a retest, identify the changed design and protocol revisions and link the earlier failure.
A change to value suppression in an error state may also affect other display states or transitions; assess those effects to determine further verification.
Neither “repeat everything” nor “repeat only the failed item” is appropriate as an automatic rule for every change.
The report should distinguish completed evidence from open actions and provide usable information for subsequent design reviews.

Related articles and sources

[1] MDSAP Audit Approach, AU P0002.011, revision 2026-08-03.
Chapter 1 Task 8, Chapter 5 Tasks 7 and 14, and Annex 1 on design and development files.
Audit guidance is distinct from the ISO standard and national regulations.

[2] FDA, Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions, 2019.
Nonbinding guidance for US premarket submissions, used within its stated scope.

[3] ISO 13485:2016 official information.
This explanation connects clause purposes and does not replace the full standard or the organization's controlled procedures.
Sources checked: 2026-11-01.

Continue through ISO 13485 clauses

Compare clause purposes and supporting evidence

Let's talk

Apply what you have read
to your product.

Tell us your product's current stage and the questions you need to review.

Discuss your questions ↗
MedQraft / Let's talk

Project inquiry

Start with your challenge.
We will work through it together.

Tell us your product's current stage and the support you need.
We will review the questions and define a suitable scope of support.

Contact@medqraft.co.kr ↗
Information: name, company, email, inquiry and service of interest · Purpose: responding to your inquiry · Retention: until the inquiry is resolved Read the privacy policy
Privacy policy

MedQraft / Privacy

Privacy policy

MedQraft processes only the information needed for website inquiries.
We delete the collected inquiry information when the inquiry is resolved.

Effective date: September 18, 2026

1. Scope and purposes of processing

This policy applies to visits and business inquiries at www.medqraft.co.kr.
We process personal information to review and respond to inquiries, discuss the scope of support, and operate the website reliably.
Customer business data stored in Q-Atelier is governed by the relevant service agreement and separate notices.

2. Information collected and retention

Required information
Name, email address and inquiry
Optional information
Company, selected service of interest, and contact or business details you provide in your inquiry
Collection methods
Website inquiry forms and email inquiries
Basis for processing
Consent to collection and use provided through the inquiry form.
Where necessary to fulfill a request relating to the formation or performance of a contract, we process information on the applicable legal basis.
Retention period
We retain inquiry information until the inquiry is resolved, then delete it without delay.
If information becomes part of a separate contract, we separately explain the required information and retention period.
Where retention is legally required, only the relevant information is stored separately.

You may decline to provide required information, but this will prevent submission of the website inquiry form.
You can make an inquiry without providing optional information.
Do not submit sensitive information unnecessary for the inquiry, such as national identification numbers or patient information.

3. Service providers and external services

Inquiries are forwarded from the website server through an email delivery service to our company mailbox.
We do not maintain a separate inquiry database in this processing workflow.

Vercel Inc.
Website hosting, server-side inquiry processing, operations and security.
Technical information such as IP address, request time, browser and request details may be processed.
Plus Five Five, Inc. (Resend)
Delivery of inquiry notification emails containing name, email, company, service of interest and inquiry details, and management of delivery logs
Gabia Hiworks
Receipt, storage and replies to inquiries in our company mailbox

We do not sell inquiry information or provide it for advertising purposes.
We do not disclose it for an independent third party's purposes unless there is a lawful basis, such as a legal requirement or separate consent.

4. Overseas processing and storage

Overseas cloud services are used to process website requests and deliver emails.
Information entered in an inquiry is transmitted over an encrypted connection upon submission.

Vercel Inc. / United States
Information is transmitted for inquiry request processing and hosting.
Privacy contact: privacy@vercel.com.
Operational records are subject to the retention periods of the services and settings used and Vercel's processing policies.
Vercel data processing information
Plus Five Five, Inc. (Resend) / United States
Information is transmitted upon submission to deliver the inquiry email and manage delivery records.
Privacy contact: privacy@resend.com.
Resend's published default retention period for emails and logs on its standard plans is 30 days.
Earlier deletion may be requested from the provider.
Resend retention and deletion information

Deleting inquiry emails from our mailbox and deleting technical records held by external providers involve different processes.
You may contact the person listed below about deletion, restriction of processing or overseas processing.
If you do not wish to use overseas processing through the website form, you may email us directly.
The processing policies of the email services used will still apply.

5. Deletion and safeguards

After confirming that an inquiry has been resolved, we delete the email and separately saved copies.
Printed materials, if any, are destroyed by shredding or another method that prevents reuse.
Where deletion of information held by an external processor is necessary, we submit a request through that provider's deletion procedure.

Access to personal information is restricted to staff who need it for their work.
We protect information through HTTPS transmission and account access controls.
We do not publish materials containing personal information in public repositories or on public pages.

6. Your rights

You may request access, correction, deletion, restriction of processing or withdrawal of consent by contacting the designated email address.
We will verify your identity to the extent necessary, process the request under applicable law, and notify you of the result.
If we cannot fulfill a request, for example because of a legal retention obligation, we will explain the reason.
For requests made through a representative, we may verify their authority.

7. Access information and external content

We use Naver conversion tracking to measure visits and completed inquiries attributable to Naver search advertising.
Technical information such as the visited URL, referring page, browser and cookies, and whether an inquiry was completed, may be processed.
The name, email, company and inquiry text entered in the form are not included in conversion events.
You can block or delete cookies in your browser.

Server access information may be processed for website delivery and security.
Vercel Web Analytics is used for anonymous visitor statistics on the introduction page and Insights.
Insights statistics aggregate article views, referral sources, countries and device types.
Visitor analytics do not use analytics cookies.
Names, emails, companies and inquiry text from the consultation form are not included in analytics events.
Instagram-specific links use UTM parameters to identify referral sources.
Loading external video or fonts on the website may transmit technical information, including IP address and browser details, to the relevant provider.

You can block or delete cookies and site data in your browser settings.
Some features or external content may be restricted depending on your settings.

8. Privacy officer and contact

Business
MedQraft / Representative: Byeonghwi Kim
Privacy officer and inquiry contact
Byeonghwi Kim
Email
Contact@medqraft.co.kr
Postal inquiries
Suite 222, 56 Geumto-ro 80beon-gil, Sujeong-gu, Seongnam-si, Gyeonggi-do, Republic of Korea (Geumto-dong, Widen Tower)

For privacy-related complaints, you may contact the Personal Information Infringement Report Center or the Personal Information Dispute Mediation Committee for assistance.

9. Changes to this policy

Changes to processing purposes, information collected, retention periods or service providers will be published on this page with their effective date.

September 18, 2026: Website inquiry privacy policy first published

October 14, 2026: Added information on Insights visitor analytics and Instagram referral measurement