MedQraft
← All insights
QMS / ISO 13485

ISO 13485 7.3.7 design validation: intended use, plans, records and reports

Understand verification versus validation and connect intended use to evaluation plans, execution records, reports and controlled design evidence.

Last regulatory review: 2026-11-02 Rev. 1.0 14 min read

Design verification establishes whether design outputs meet design inputs.
Design validation establishes whether the product meets requirements for its specified application or intended use.
Linking the two requires more than a passed input test: the evidence must also address who will use the product, in which environment and for which purpose.

This article discusses ISO 13485:2016 clauses 7.3.6 and 7.3.7 through a fictional thermometer example.
Document titles and identifiers below are MedQraft implementation examples, not prescribed forms.

1. Different reference points for the conclusion

Aspect Design verification, 7.3.6 Design validation, 7.3.7
Reference point Design input requirements Requirements for the specified application or intended use
Example question Does the previous reading disappear in the specified error state? Can intended users accomplish the measurement purpose in the relevant situation?
Possible evidence Display tests, analysis and inspection Appropriate performance, use-related and applicable clinical evidence
Scope of conclusion Requirements and configurations actually examined Intended use, populations, environments and configurations evaluated

The distinction is not simply machine testing versus testing with people.
Usability-related inputs can be verified, and validation can involve several forms of technical and clinical evidence.
Identify the requirement and the conclusion each activity supports, rather than classifying it solely by its location or method.

Verification connects input requirements and outputs; validation connects intended use and product evidence
The two activities assess evidence against different reference points.
A shared data source still needs a clear link to each requirement and conclusion.

2. A cleared reading does not establish completion of measurement

Consider a fictional home-use thermometer with an input requiring the previous reading to disappear in a specified error state.
A test confirms that behavior.
The result supports that display requirement, rather than every requirement for the product.

During a simulated home-use evaluation, a participant sees the blank screen and puts the thermometer away.
In an interview after the task, the participant explains that the blank screen appeared to mean the power was off.
This observation provides evidence about interpretation and continuation of measurement.
One observation does not establish a population failure rate or a conclusion about the entire device.

The next step is to investigate how the display state, error message, repeat-measurement interaction and instructions work together.
Adding text is a design option to evaluate, not proof of resolution.
After a change, determine the appropriate scope of verification and validation from its impact.

This is a development discovery and improvement example, not a claim of completed final validation.
It addresses one use-related aspect.
Measurement performance and other aspects of the intended use require their own consideration.

3. Define the scope before writing the protocol

Scope element Useful planning information Consequence of omission
Intended users Relevant abilities, experience and user groups Developer familiarity is mistaken for user understanding
Use environment Location, relevant lighting or noise, available assistance Evaluation conditions omit important constraints
Product configuration Model, hardware and software versions, accessories, labeling and instructions Results cannot be associated with a particular configuration
Purpose and tasks Intended outcome, use sequence and risk-related tasks Preference feedback replaces evidence about essential use
Representativeness Relationship between evaluated and final products, differences and their impact Early prototype results are applied to a changed final product without justification

Validation needs a planned approach and a defensible relationship between the evaluated product and the conclusion.
Evidence supporting intended use and completion before release are important audit considerations.
See the official MDSAP Audit Approach, Chapter 5 Task 10.

4. Connect the plan, execution record and report

Producing three documents is less important than maintaining a common evaluation identity across them.
Product versions, activity identifiers and the source of acceptance criteria make the conclusion traceable to evidence.

Example document Timing and purpose Suggested contents
Validation plan DVAL-P01 Before execution: establish scope and responsibility Intended use, configuration, activities, schedule, responsibilities, related risks, applicable references and rationale for sampling and analysis
Protocol DVAL-T01 Before execution: define methods and decisions Tasks, starting conditions, information provided, observations, predetermined criteria, stopping and deviation rules, raw-data collection
Execution record DVAL-RUN01 During execution: preserve what actually occurred Date and personnel, versions, coded participant identity, actual behavior, equipment and environment, assistance, deviations and raw-data locations
Results report DVAL-R01 After execution: compare evidence with criteria Planned versus completed scope, item-level outcomes, source references, unmet or unperformed items, deviation impact, limitations and actions
Traceability matrix TRACE-01 Maintained throughout development and changes User need → risk → evaluation item → execution record → conclusion → change and reassessment

A plan and protocol may be combined, and a report may contain the traceability matrix.
Choose a structure appropriate to the organization's procedures and product complexity.
Consolidating documents should preserve the information needed to reconstruct decisions.

Plan P01 and item T01 connect execution record RUN01 with report R01
The report references both the planned criteria and the actual record.
Stable activity identifiers help reviewers locate the basis of the conclusion.
Identifiers are illustrative.

5. Follow one evaluation item through the evidence chain

The following example illustrates traceability.
It is neither a product acceptance standard nor an actual study result.

Connection Illustrative entry
User need NEED-01 Intended home users can understand the measurement sequence and complete temperature measurement
Risk review RISK-07 Examine misinterpretation of an error state, failure to repeat measurement and possible resulting harm
Evaluation item T-01 Observe error interpretation and subsequent behavior within the specified scenario
Protocol Predetermine starting conditions, information provided, observations, post-task questions and how assistance is recorded
Execution RUN-01 Participant puts the device away after the screen becomes blank; subsequently reports believing it had powered off
Report R-01 Distinguish observed behavior from reported interpretation; assess against the predetermined criteria and connect investigation to actions
Change and reassessment Link change ID, revised configuration, impact assessment and the rationale and results of further verification or validation

Avoid recording only an interpretation such as “the user was confused.”
Record the action, its timing and the participant's response to a stated question separately.
If assistance was provided, preserve that fact and evaluate its effect on the result.
Distinguishing post-task interviews from intervention during the task helps protect the meaning of the observation.

Acceptance criteria should be established before results are known.
Do not introduce a convenient threshold retrospectively.
Select participant or sample numbers and repetitions using the evaluation purpose, risk, method and applicable guidance.
Address applicable statistical methods and the rationale for sample size in planning; a single universal number is inappropriate.

6. Assign preparation, review and approval roles

A standard clause should not be treated as a fixed department signature box.
Establish authority and responsibility through organizational procedures and the development plan, with relevant competence involved in review.
One possible allocation is:

  • Preparation: Development and evaluation personnel translate the intended use and configuration into the plan and protocol.
    Personnel performing the activity record observations and deviations.
  • Technical review: Appropriate reviewers assess whether the methods address the purpose, risk-related items, representativeness and scope of conclusions.
    Usability, clinical or statistical expertise may be needed.
  • Quality and regulatory review: Reviewers check applicable requirements, revision control, traceability, unresolved issues and approval conditions.
    Department names may differ across organizations.
  • Approval: Authorized personnel approve use of the plan or acceptance of the report under the applicable procedure.
    A signature does not supply missing evidence.

The report should distinguish completion of an activity from sufficient evidence for the intended purpose.
Identify unperformed work, unresolved issues and the responsibility for follow-up.

7. Standards and document-control connections

Reference Role in this work Practical connection
ISO 13485:2016, 7.3.7 Design and development validation Intended use, planned methods and criteria, results and necessary actions
ISO 13485:2016, 4.2.4 Document control Review, approval, revision identification and controlled availability of plans and protocols
ISO 13485:2016, 4.2.5 Record control Identification, protection, retrieval and retention of execution records and results
ISO 13485:2016, 7.3.9 Design and development changes Impact of display, behavior or instruction changes and the scope of reassessment
ISO 13485:2016, 7.3.10 Design and development files Inclusion of, or references to, relevant evidence for the product
ISO 14971:2019 Medical device risk management Use-related situations, risk controls and evidence of their effectiveness
IEC 62366-1:2015 with AMD1:2020 Safety-related usability engineering Use-related risk processes; not a replacement for all design validation

ISO 13485 is not a product test method that assigns a universal number of users.
Identify product-specific performance and safety standards and market requirements separately.
Record the edition and applicability of each reference used to justify the evaluation.

The FDA guidance Applying Human Factors and Usability Engineering to Medical Devices gives recommendations concerning intended users, uses, environments and use-related risk reduction.
On the review date, the official page identifies an August 2026 version.
Its recommendations should not be presented as mandatory document templates for every jurisdiction.
FDA guidance page

Where clinical evaluation is applicable, it does not automatically mean that a new clinical trial is always required.
Determine the evidence needed for the device and regulatory context.
Completion of a usability evaluation must not be generalized into proof of clinical effectiveness.

8. Revise documents while preserving the history of records

When revising a plan, identify the reason, effective timing, affected activities and approval status.
Do not make an earlier activity appear to have been conducted under a later protocol.
Preserve the connection between the previous plan and its execution records, and identify the revision used for additional work.

Record corrections should follow the organization's procedure and preserve traceability of the original information and correction.
Reissuing a report should retain links to raw data, the earlier conclusion, the reason for revision and the revised conclusion.
Where participant information is involved, define access restrictions and privacy arrangements.

Do not assign an arbitrary period such as five years to every record.
Determine retention requirements using device lifetime, applicable regulation and contractual conditions within the organization's record-control process.
For electronic evidence, consider retrieval, permissions, backup, recovery and change history as well as storage location.

Previous evidence remains linked to change assessment, revised evaluation and the design and development file
A favorable new result does not overwrite the earlier evidence.
Keep the previous configuration, change decision and new evaluation traceable together.

9. Review the links before approving the conclusion

A reviewer should be able to identify the product configuration and intended use from the report, then move from its conclusion to predetermined criteria, actual records and raw data.
Explain how deviations and unresolved items affect the conclusion.

For the fictional thermometer, “the error display was changed” is only an action status.
The evidence should show which use problem was addressed and how the revised configuration was assessed against the relevant requirements and intended purpose.
Include or reference that chain in the design and development file so it remains useful in subsequent changes.

Related insights

Sources and review scope

Reviewed 2026-11-02 using official standards metadata and scope information, the MDSAP audit approach and FDA guidance information.
This article does not reproduce standard text or forms.
For implementation, consult the organization's controlled editions and applicable regulatory texts.

  1. ISO 13485:2016, medical device quality management systems.
  2. MDSAP Audit Approach AU P0002.011, 2026-08-03, document and record controls and design/development audit considerations.
  3. ISO 14971:2019, medical device risk management.
  4. IEC 62366-1:2015 AMD1:2020, safety-related usability engineering.
  5. FDA human factors and usability engineering guidance, official page labeled August 2026.

Continue through ISO 13485 clauses

Compare clause purposes and supporting evidence

Let's talk

Apply what you have read
to your product.

Tell us your product's current stage and the questions you need to review.

Discuss your questions ↗
MedQraft / Let's talk

Project inquiry

Start with your challenge.
We will work through it together.

Tell us your product's current stage and the support you need.
We will review the questions and define a suitable scope of support.

Contact@medqraft.co.kr ↗
Information: name, company, email, inquiry and service of interest · Purpose: responding to your inquiry · Retention: until the inquiry is resolved Read the privacy policy
Privacy policy

MedQraft / Privacy

Privacy policy

MedQraft processes only the information needed for website inquiries.
We delete the collected inquiry information when the inquiry is resolved.

Effective date: September 18, 2026

1. Scope and purposes of processing

This policy applies to visits and business inquiries at www.medqraft.co.kr.
We process personal information to review and respond to inquiries, discuss the scope of support, and operate the website reliably.
Customer business data stored in Q-Atelier is governed by the relevant service agreement and separate notices.

2. Information collected and retention

Required information
Name, email address and inquiry
Optional information
Company, selected service of interest, and contact or business details you provide in your inquiry
Collection methods
Website inquiry forms and email inquiries
Basis for processing
Consent to collection and use provided through the inquiry form.
Where necessary to fulfill a request relating to the formation or performance of a contract, we process information on the applicable legal basis.
Retention period
We retain inquiry information until the inquiry is resolved, then delete it without delay.
If information becomes part of a separate contract, we separately explain the required information and retention period.
Where retention is legally required, only the relevant information is stored separately.

You may decline to provide required information, but this will prevent submission of the website inquiry form.
You can make an inquiry without providing optional information.
Do not submit sensitive information unnecessary for the inquiry, such as national identification numbers or patient information.

3. Service providers and external services

Inquiries are forwarded from the website server through an email delivery service to our company mailbox.
We do not maintain a separate inquiry database in this processing workflow.

Vercel Inc.
Website hosting, server-side inquiry processing, operations and security.
Technical information such as IP address, request time, browser and request details may be processed.
Plus Five Five, Inc. (Resend)
Delivery of inquiry notification emails containing name, email, company, service of interest and inquiry details, and management of delivery logs
Gabia Hiworks
Receipt, storage and replies to inquiries in our company mailbox

We do not sell inquiry information or provide it for advertising purposes.
We do not disclose it for an independent third party's purposes unless there is a lawful basis, such as a legal requirement or separate consent.

4. Overseas processing and storage

Overseas cloud services are used to process website requests and deliver emails.
Information entered in an inquiry is transmitted over an encrypted connection upon submission.

Vercel Inc. / United States
Information is transmitted for inquiry request processing and hosting.
Privacy contact: privacy@vercel.com.
Operational records are subject to the retention periods of the services and settings used and Vercel's processing policies.
Vercel data processing information
Plus Five Five, Inc. (Resend) / United States
Information is transmitted upon submission to deliver the inquiry email and manage delivery records.
Privacy contact: privacy@resend.com.
Resend's published default retention period for emails and logs on its standard plans is 30 days.
Earlier deletion may be requested from the provider.
Resend retention and deletion information

Deleting inquiry emails from our mailbox and deleting technical records held by external providers involve different processes.
You may contact the person listed below about deletion, restriction of processing or overseas processing.
If you do not wish to use overseas processing through the website form, you may email us directly.
The processing policies of the email services used will still apply.

5. Deletion and safeguards

After confirming that an inquiry has been resolved, we delete the email and separately saved copies.
Printed materials, if any, are destroyed by shredding or another method that prevents reuse.
Where deletion of information held by an external processor is necessary, we submit a request through that provider's deletion procedure.

Access to personal information is restricted to staff who need it for their work.
We protect information through HTTPS transmission and account access controls.
We do not publish materials containing personal information in public repositories or on public pages.

6. Your rights

You may request access, correction, deletion, restriction of processing or withdrawal of consent by contacting the designated email address.
We will verify your identity to the extent necessary, process the request under applicable law, and notify you of the result.
If we cannot fulfill a request, for example because of a legal retention obligation, we will explain the reason.
For requests made through a representative, we may verify their authority.

7. Access information and external content

We use Naver conversion tracking to measure visits and completed inquiries attributable to Naver search advertising.
Technical information such as the visited URL, referring page, browser and cookies, and whether an inquiry was completed, may be processed.
The name, email, company and inquiry text entered in the form are not included in conversion events.
You can block or delete cookies in your browser.

Server access information may be processed for website delivery and security.
Vercel Web Analytics is used for anonymous visitor statistics on the introduction page and Insights.
Insights statistics aggregate article views, referral sources, countries and device types.
Visitor analytics do not use analytics cookies.
Names, emails, companies and inquiry text from the consultation form are not included in analytics events.
Instagram-specific links use UTM parameters to identify referral sources.
Loading external video or fonts on the website may transmit technical information, including IP address and browser details, to the relevant provider.

You can block or delete cookies and site data in your browser settings.
Some features or external content may be restricted depending on your settings.

8. Privacy officer and contact

Business
MedQraft / Representative: Byeonghwi Kim
Privacy officer and inquiry contact
Byeonghwi Kim
Email
Contact@medqraft.co.kr
Postal inquiries
Suite 222, 56 Geumto-ro 80beon-gil, Sujeong-gu, Seongnam-si, Gyeonggi-do, Republic of Korea (Geumto-dong, Widen Tower)

For privacy-related complaints, you may contact the Personal Information Infringement Report Center or the Personal Information Dispute Mediation Committee for assistance.

9. Changes to this policy

Changes to processing purposes, information collected, retention periods or service providers will be published on this page with their effective date.

September 18, 2026: Website inquiry privacy policy first published

October 14, 2026: Added information on Insights visitor analytics and Instagram referral measurement